This transducer is essentially a convection-enhanced Pirani gauge and associated electronics integrated into a single unit. It requires DC power to operate. It provides an analog output as well as two setpoint relays. It provides no direct display of pressure, but has LED indicators to show that DC power is present and to show the status of the set point relays
- MKS HPS type 907 Analog Convection Transducer
- part number 109070019CE (CE marked version of the 109070019)
- nonlinear analog output
- two process control set point relays
- KF-16 flange
- mating electrical connector included
- measuring range (for air/nitrogen):
- 1.0 x 10-3 to 1.0 x 103torr
- 1.3 x 10-3 to 1.3 x 103mbar
- 1.3 x 10-1 to 1.3 x 105 Pa
- useful set point range (for air/nitrogen):
- 5.0 x 10-3 to 1.0 x 103torr
- 6.5 x 10-3 to 1.3 x 103mbar
- 6.5 x 10-1 to 1.3 x 105 Pa
- operating temperature range: 0 °C to 40 °C
- maximum bakeout temperature: 100 °C (with electronics removed from sensor)
- installation orientation: horizontal only
- input power requirements: 12 to 30 VDC, 0.2 A
- analog output: 0.38 to 5.66 VDC, nonlinear; 1 k ohm maximum output impedance
- relay characteristics:
- SPDT, rated for 1 A @ 24 VAC/DC, resistive load
- relay response 150 msec maximum