- Penning Cold Cathode Sensor
- Range: (10-3 to 10-7 Torr)
- Calibration Medium: Dry air
- Available end connections
(NPT, 3/4" and 1-1/4" tube, NW25) - Rugged Design: Excellent for use in contaminatedenvironments
- Cleanable for like-new performance
- No recalibration is needed after cleaning.
The 7B cold cathode is a penning-type sensor. In this configurationthe anode is placed midway between two parallel-connected cathodes.The anode is a loop of metal wire whose plane is parallel to thatof the cathodes. A potential difference of a few kilovolts ismaintained between the anode and the cathodes. In addition, amagnetic field is applied between the cathodes by a permanentmagnet, usually external of the gauge body. Electrons emitted fromeither of the two cathodes must travel in helical paths due to themagnetic field, eventually reaching the anode, which carries a highpositive charge. During the course of this long electron path, manyof the electrons collide with the molecules of residual gas,creating positive ions, which travel more directly to the cathodes.The ionization current is then converted to a vacuum reading.
Penning冷阴极传感器
范围:(10-3到10-7托)
校准介质:干燥的空气
可用的端部连接(“NPT”,3/4“和1-1/4”管,NW25)
坚固的设计:非常适用于在污染环境
可清洁如新的性能
清洗后无需重新校准
7B Cold Cathode Penning Gauges
(10-3 to 10-7 Torr)GAUGE TUBE PHIL BRS 3/4-NPT 2-2100-263 GAUGE TUBE PHIL BRS 1-1/4" ST 2-2100-266 GAUGE TUBE PHIL BRS 3/4" ST 2-2100-264 GAUGE TUBE PHIL SS NW25 FLANGE 2-2100-272
美国Televac 7B2冷规